Abstract

Apparatus is described for measurement of sound speed and ultrasound attenuation coefficients by the substitution technique in the frequency range 3 to 8 MHz. Phase-cancellation artifacts leading to overestimation of attenuation coefficients are avoided by use of an acoustoelectric transducer. Specimens confined by polystyrene windows can be interrogated by focused ultrasound beams at selected locations spaced on a grid of 3 × 3 mm voxels. Pulse time of flight is measured with an accuracy of 30 ns, yielding sound speeds accurate to ±6.7 m/s, for samples 10 mm thick. Uncertainties in measured insertion losses range from 0.1 dB in low-loss (10 dB) specimens to 0.5 dB in high-loss (25 dB) specimens.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.