Abstract

In response to the issues of insufficient stability and accuracy in dry chemical detection using reflectance photometry, caused by the divergence and multiple internal reflections of the reflected light signal from the sample and the multilayer dry film test strip, a dry chemical reflectance photometry detection system based on an integrating sphere is designed. Firstly, an integrating sphere device is incorporated to reduce signal divergence and loss, ensuring even detection of the sample's reflected light signal and improving detection stability. Secondly, Light Tools optical simulation analysis is performed, and an integrating sphere detection model is established. Thirdly, the Williams-Clapper equation is employed to correct the error in reflectance density caused by multiple internal reflections, enhancing detection accuracy. Experimental validation demonstrates that the developed integrating sphere-based dry chemical reflectance photometry detection system improves the stability and accuracy of the detection system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.