Abstract

The identification and real time detection of explosives and hazardous materials are of great interest to the Army and environmental monitoring/protection agencies. The application and efficiency of the remote Raman spectroscopy system for real time detection and identification of explosives and other hazardous chemicals of interest, air pollution monitoring, planetary and geological mineral analysis at various standoff distances have been demonstrated. In this paper, we report the adequacy of stand-off Raman system for remote detection and identification of chemicals in water using dissolved sodium nitrate and ammonium nitrate for concentrations between 200ppm and 5000ppm. Nitrates are used in explosives and are also necessary nutrients required for effective fertilizers. The nitrates in fertilizers are considered as potential sources of atmospheric and water pollution. The standoff Raman system used in this work consists of a 2-inch refracting telescope for collecting the scattered Raman light and a 785nm laser operating at 400mW coupled with a small portable spectrometer.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call