Abstract
A flow-injection differential-pulse anodic stripping voltammetry (d.p.a.s.v.) method is modified so that interferences from humic acids or surfactants are eliminated. The injected, slightly acidic sample is passed through a silica anion-exchanger column to remove compoundswith a strong tendency to adsorb to the electrode. The sample then passes to a chelate ion-exchange column containing immobilized 8-quinolinol. The metal ions are retained and later eluted with acid into the voltammetric cell. The results show that the interferences from up to 500 mg 1 –1 humic acid or at least 50 mg 1 -1 Triton X-100 can be removed and that the metal ion can be determined in a range similar to that for normal d.p.a.s.v. methods. The complete cycle time for a determination was 12 min.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.