Abstract
A gas chromatographic method has been developed for the determination of amitriptyline and nortriptyline in plasma. OV-17 is used in a 1 m long packed column, with a flame ionization detector and an electronic integrator. Five internal standards are added. The base-specific extraction procedure and the method of calibrating the chromatograph are described in detail. The accuracy, precision and reliability of the method are demonstrated by the results of nearly 700 determinations of each drug, at concentrations ranging from 5 to 400 ng/ml in the plasma. An interlaboratory comparison with a double radioactive isotope derivative assay for nortriptyline has also shown satisfactory agreement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.