Abstract

[Purpose] This study investigated the reliability of the Special Support System. [Subjects and Methods] Eighteen female and 14 male volunteers participated in this study. Participants were asked to sit on the Special Support System with their chins tucked in, spines straight, pelvis neutrally positioned, and their hands placed on their thighs. They were also asked to flex their hips, knees, and ankles to approximately 90 degrees and to put their feet flat on the floor. The total contact area, mean total pressure, as well as mean and peak pressures of each quadrant were each measured 15 times. Test-retest reliability was analyzed for inflated air pressure, and pressure redistribution values by using intraclass correlation coefficients. [Results] The intraclass correlation coefficient was greater than 0.89 for inflated air pressure and greater than 0.92 for total contact area, mean total pressure, and each quadrant’s mean and peak pressure. [Conclusion] The findings suggest that the Special Support System is reliable and can be used as an alternative method for redistributing sitting pressure.