Abstract
[Purpose] Quadriceps muscle strength is essential for daily living activities. Therefore, we developed a compact and simple lower limb muscle strength measuring device (LocomoScan [LCS]). This study aimed to compare LCS with other instruments to analyze its simplicity, reproducibility, and accuracy. [Participants and Methods] One hundred and four healthy university students (56 males and 48 females) were included in the study. The knee extension force was measured using LCS, and the knee extension torque was measured using other devices (Cybex). In addition, lower leg muscle mass was measured using a body composition meter. The reproducibility of LCS and the correlation between the knee extension torque and lower leg muscle mass were evaluated. [Results] The measurement reproducibility of LCS was significantly higher. The knee extension force confirmed the proportional relative reliability of Cybex with knee extension torque. A relationship between knee extension force and lower limb muscle mass was also observed, indicating that muscle mass cannot be estimated as muscle strength. [Conclusion] The high reproducibility of the knee extension force measurement using LCS demonstrates its potential as a portable alternative instrument for muscle strength measurement in clinical practice. Therefore, LCS device is a simple and effective tool for assessing muscle strength.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.