Abstract
<p>The important things on Leksell Gamma Knife (LGK) is determination of Relative Output Factor (ROF), which is required for beam data collection (BDC) process, as an input value to the Treatment Planning System (TPS). The ROF is also needed to traceability of dosimetry measurements and patient safety. ROF measurements on LGK using PTW PinPoint 3D ionization chamber: 0.016 cm<sup>3</sup>, PTW Semiflex: 0.125 cm<sup>3</sup>, and Exradin A16: 0.017 cm<sup>3</sup> at collimators 4, 8 and 16 mm. and gafchromic films. ROF values obtained in hospital-A, using PTW PinPoint 3D was: [(1.00 ± 0.002)], [(0.80 ± 0.02]); [(0.42 ± 0.002)]; and using Exradin A16 [(1.00 ± 0]; [(0, 87 ± 0)]; [(0.69 ± 0.0002)]. And using gafchromic films was: [(1.00 ± 0.006]; (0.96 ± 0.005); (0.86 ± 0.06)]; respectively for collimators 4, 8 and 16 mm, and Measurement of ROF in hospital-B using using a PTW PinPoint 3D was: [(1 ± 0); (0.75 ± 0); and (0.33 ± 0)] and using Exradin A16 was: [(1; 0.87; and 0.69)] and gafchromic films was: [(1.00 ± 0.03); (0.98 ± 0.045); and (0.85 ± 0.09)]; The results of ROF’s measurement were compared with the ROF values of Novotny, et al (2009) and TPS. The Gafchromic’s ROF values were relatively closer than both of them, while the ROF value of ionization. chamber at 4 mm collimators was relatively lower, because measurement in a small field occurs an averaging effect volume and lack of lateral charge particle equilibrium (LCPE).</p>
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.