Abstract

PurposeTo compare the intensity of muscle contractions in irreversible electroporation (IRE) treatments when traditional IRE and high-frequency IRE (H-FIRE) waveforms are used in combination with a single applicator and distal grounding pad (A+GP) configuration. Materials and MethodsAn ex vivo in situ porcine model was used to compare muscle contractions induced by traditional monopolar IRE waveforms vs high-frequency bipolar IRE waveforms. Pulses with voltages between 200 and 5,000 V were investigated, and muscle contractions were recorded by using accelerometers placed on or near the applicators. ResultsH-FIRE waveforms reduced the intensity of muscle contractions in comparison with traditional monopolar IRE pulses. A high-energy burst of 2-μs alternating-polarity pulses energized for 200 μs at 4,500 V produced less intense muscle contractions than traditional IRE pulses, which were 25–100 μs in duration at 3,000 V. ConclusionsH-FIRE appears to be an effective technique to mitigate the muscle contractions associated with traditional IRE pulses. This may enable the use of voltages greater than 3,000 V necessary for the creation of large ablations in vivo.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.