Abstract

A rapid and specific high-performance liquid chromatographic (HPLC) method has been developed for quantitative analysis of famotidine in pharmaceutical dosage forms. The method was applied to medicines on the Turkish market. A LiChrospher RP-18 column, HP-1050 UV-detector and methanol—0·1 M aqueous ammonium acetate (30 : 70) as the solvent system (flow rate 1·0 mL min−1) were used. Quantitative determination of famotidine was carried out at 254 nm wavelength. The method was found to be rapid—under 5 min per assay—and was selective enough to allow the determination of famotidine in the presence of certain excipients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.