Abstract
A highly sensitive and simple LC-MS/MS method after one-step protein precipitation was developed and validated for determination of pidotimod (CAS 121808-62-6) in human plasma using dextrophan (CAS 125-73-5) as internal standard (IS). Pidotimod and IS were separated on a YMC-ODS-AQ C18 column using 0.5% formic acid and methanol as a mobile phase at a flow rate of 0.3 mL/min. Detection was performed on positive ion mode of the transitions at 245.0→134.0 for pidotimod and 258.1→157.0 for IS by selected reaction monitoring (SRM). The assay exhibited a linear range of 0.05-10.0 µg/mL. The lower limit of quantification were 0.05 µg/mL. Validation results indicated that the accuracy as determined from quality control samples was in the range of - 4.00-6.48%. Intra-day and inter-day precision was ≤ 8.35% and ≤ 8.00%, respectively. The developed method was successfully applied to a bioequivalence study in 20 healthy Chinese volunteers following a single oral dose of 800 mg pidotimod. The simple, inexpensive protein precipitation and high-throughput method makes it a suitable and valuable tool in the investigation of the clinical pharmacokinetics and bioequivalence.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.