Abstract
Photoacoustic (PA) imaging technology is expected to be applied to clinical assessment for peripheral vascularity. We started a clinical evaluation with the prototype PA imaging system we recently developed. Prototype PA imaging system was composed with in-house Q-switched Alexandrite laser system which emits short-pulsed laser with 750 nm wavelength, handheld ultrasound transducer where illumination optics were integrated and signal processing for PA image reconstruction implemented in the clinical ultrasound (US) system. For the purpose of quantitative assessment of PA images, an image analyzing function has been developed and applied to clinical PA images. In this analyzing function, vascularity derived from PA signal intensity ranged for prescribed threshold was defined as a numerical index of vessel fulfillment and calculated for the prescribed region of interest (ROI). Skin surface was automatically detected by utilizing B-mode image acquired simultaneously with PA image. Skinsurface position is utilized to place the ROI objectively while avoiding unwanted signals such as artifacts which were imposed due to melanin pigment in the epidermal layer which absorbs laser emission and generates strong PA signals. Multiple images were available to support the scanned image set for 3D viewing. PA images for several fingers of patients with systemic sclerosis (SSc) were quantitatively assessed. Since the artifact region is trimmed off in PA images, the visibility of vessels with rather low PA signal intensity on the 3D projection image was enhanced and the reliability of the quantitative analysis was improved.