Abstract
Quality control of clinical Nuclear medicine instruments” i.e. used, used in this research work i.e. radioisotope dose calibrator and SPECT gamma camera is most important for reliable operational characteristics in this research work different tests has been done taking data from “ GE- Infinia Hawkeye” SPECT Gamma camera at “DINAR” Cancer hospital D.I.Khan.Auto zero, background, linearity, geometry, constancy and accuracy tests were performed in the case of radioisotopes dose calibrator while tests performed in case of SPECT gamma camera includes background, energy peak, C.O.R and uniformity tests. In conclusion, all the obtained results were within acceptable range stated in operational manual of “GE- Infinia hawk eye” SPECT Gamma Camera and that of radioisotopes dose calibrator was in accordance with IAEA (International atomic energy agency).Thus “Radioisotopes dose calibrator” and “SPECT gamma camera” is in good working condition at DINAR cancer hospital D.I.Khan
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.