Abstract

Standard lead aprons do not protect the female breast adequately from radiation exposure, which has been associated with breast cancer in healthcare workers. A novel lead shield was designed to reduce radiation to the breast, axilla, and thyroid (BAT). A procedure room was simulated with an anthropomorphic phantom representing the operator. Dosimeters were positioned on the outer quadrant of each breast, the chest, the thyroid, and deep inside of a phantom acrylic female torso with neck and head. Standard lead vest plus a thyroid shield was used as control and compared to standard lead vest plus BAT shield. Three operator and two image receptor positions were tested. The reductions in radiation exposure were calculated. The standard vest plus BAT shield provided significant reductions in radiation exposure for all anatomic locations compared to control. When averaging all operator positions, the BAT provided reductions of 91% (p < 0.0001) for near breast. Reductions for far breast, chest, thyroid, and deep tissues were 76% (p = 0.016), 94% (p < 0.0001), 52% (p = 0.026), and 60% (p = 0.004). With operator 90° to the table using a cross-table lateral beam, the BAT provided a 97.7% reduction in radiation to the near breast and significant reduction in radiation to the chest, thyroid, and deep tissues. The BAT shield reduces radiation exposure to the breast, chest, thyroid and deep hematopoietic tissues. Such shields could benefit healthcare workers to reduce the risk of breast cancer and other radiation-associated cancers.