Abstract
The radiation exposure in the cath lab of patients, cardiologists, and nurses was measured during three consecutive periods of 8 weeks. The first 8 weeks the baseline radiation exposure was obtained. In the second period standard incidences for coronarography and frame rate were changed, without compromising the image quality of the examination. In the third period, a pelvic shield covered the lower part of the patient. This pilot quality project demonstrates that further significant reduction in radiation exposure of 37% is possible for patients. A significant reduction in radiation exposure of 53 and 62% was obtained among cardiologists and nurses working in the cath lab, even with the already diminished radiation exposure over the last years by better equipment and general radioprotection measures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.