Abstract

Irreversible electroporation (IRE) is a promising alternative therapy for the local treatment of prostate tumors. The procedure involves the direct insertion of needle electrodes into the target zone, and subsequent delivery of short but high-voltage pulses. Successful outcomes rely on adequate exposure of the tumor to a threshold electrical field. To aid in predicting this exposure, computational models have been developed, yet often do not incorporate the appropriate tissue-specific properties. This work aims to quantify electrical conductivity behavior during IRE for three types of tissue present in the target area of a prostate cancer ablation: the tumor tissue itself, the surrounding healthy tissue, and potential areas of necrosis within the tumor. Animal tissues were used as a stand-in for primary samples. The patient-derived prostate tumor tissue showed very similar responses to healthy porcine prostate tissue. An examination of necrotic tissue inside the tumors revealed a large difference, however, and a computational model showed that a necrotic core with differing electrical properties can cause unexpected inhomogeneities within the treatment region.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call