Abstract

In the past eight years somewhat over 3,000 patients have been examined with the electroretinogram in our department. In this period of time there have been two instances of corneal burn due to electrical energy from the recording device being fed into the patient. In neither of these instances was there any permanent damage done, but considerable discomfort on the part of the patient and concern on the part of the examiner was engendered. In an attempt to eliminate, as much as possible, this type of accident, a study of the possible sources of such technical failures was begun and a search for a means of prevention of repetition. The electrical circuit normally in use for clinical electroretinography is shown in Figure 1<i>A</i>. The patient, who is grounded, is connected by way of a contact-lens electrode to the input of an amplifier. This amplifier develops within it "stray capacities"

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.