Abstract

This work represents the first attempt to evaluate the precision of the specific oxygen uptake rate method expressed in terms of repeatability ( r) and reproducibility limits ( R). Three laboratories were involved in an inter-laboratory test for the validation of respiration analyses on six biomass samples (three composts and three biostabilized products) having different degrees of biological stability. Both the maximum specific oxygen uptake rate peak (SOUR) and the cumulative oxygen demand after 12 h (OD 12) and 20 h (OD 20) of respiration test were investigated. Precisions expressed as the relative standard deviation were in the range of 9–41%. Linear regressions found for r and R, versus OD 12 and OD 20, enabled derivation of precision values ( r and R) for all respirometric levels within the operating range. The OD 12 and OD 20 indices were found to be more adequate to indicate biological stability since they were less influenced by random errors than the SOUR index.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.