Abstract

We analysed the relationship between serum metal ions, radiological periprosthetic osteolysis and the clinical features in a series of 25 patients treated with fourth-generation metal-on-polyethylene total wrist prostheses. The mean implant follow-up was 7 years. Our results show that titanium was the main elevated serum metal ion in patients with the prostheses that we used; elevated serum cobalt or chromium values were infrequent. The risk of loosening was higher in an implant older than 6 years, with more than five periprosthetic osteolysis points according to our radiograph zone system, and serum titanium values between 26 to 31 µg/L. The presence of metallosis pseudotumours does not guarantee but increases the risk of implant failure, which may be asymptomatic or associated with little pain.Level of evidence: IV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.