The increasing demand for total hip arthroplasty (THA) in relatively young, high-demand patients has led to the use of hard-on-hard bearing surfaces. Adverse local tissue reaction/pseudotumor and elevated serum metal ion levels are commonly reported complications encountered in metal-on-metal THA, while audible articulation and rim fracture are reported in ceramic-on-ceramic THA. For this reason, ceramic-on-metal THA was implemented as an ideal hard-on-hard bearing combination. In this report, we describe a case of bilateral simultaneous ceramic-on-metal THA in a 69-year-old woman who presented 7 years postoperatively with unilateral hip pain associated with underlying pseudotumor and elevated serum cobalt and chromium ion levels. Pre-revision workup, intraoperative findings, and postoperative evaluation are included and suggest acetabular malposition as a potential source for complication.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.