Abstract
Total wrist arthroplasty (TWA) aims to restore pain-free motion to diseased joints. One such TWA, the Motec, has demonstrated good results with acceptable complication rates. It has recently been suggested that the metal-on-carbon fibre reinforced poly ether ether ketone (Mo-CFR-PEEK) version of the Motec TWA be implanted instead of the metal-on-metal version. An explant analysis was undertaken on seven Motec Mo-CFR-PEEK TWAs, revised for a variety of reasons, after a mean time of 2 years in vivo. Compared to a new Motec implant, five of the explanted metal heads and three of the CFR-PEEK cups became smoother in vivo, suggesting self-polishing and negative skewness, indicating some material loss in vivo. Two explanted cups showed indentation marks on their rims and one of these was from component impingement with embedded metallic debris. In the short-term, the articulating surfaces of Motec Mo-CFR-PEEK TWAs did not show major damage.Level of evidence: IV.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.