Abstract
AbstractThe results of an investigation into molecular weight and resolution calibration of a gel permeation chromatograph are reported. Effects of sample amounts and solvent flow rate were observed. Severe skewing and tailing of the standard chromatograms were observed at 2.0 ml/min flow rate, relative to 1.0 and 3.0 ml/min. Resolution calibration by flow reversal was inadequate when skewing and tailing or sample impurity effects were significant. Differential molecular weight distributions (MWD) of broad‐distribution polystyrenes were compared for four methods of resolution correction. Inconsistent oscillations were observed in the MWD's at low resolutions and higher molecular weights. None of the methods was completely adequate in accounting for skewing and tailing.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.