Abstract

Herein are reported improved methods of molecular weight calibration where simultaneously, peak broadening parameters (σ) are obtained through the use of multiple polydisperse molecular weight standards. There are two basic methods covered. The first and most reliable method employs the universal molecular weight calibration curve obtained using narrow MWD polystyrene standards. The second method assumes that the molecular weight calibration curve is linear on a semilog plot and should be employed where universal calibration is not practical as with aqueous SEC. Several variants of these methods involving different molecular weight data for the standards are discussed. The proposed methods have been evaluated using aqueous SEC and polydextran standards and nonaqueous SEC with polyvinylchloride standards. Previous methods of molecular weight calibration using broad MWD standards were of three basic types. Those which employ a broad MWD standard with known molecular weight distribution (1,2,3, 4,5). Those which employ one or more broad

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.