Abstract
A convenient piezoelectric driven Kelvin probe for the measurement of work function changes is described. The probe has a simple construction, is small, bakeable, and can be mounted on any 35-mm-i.d. UHV flange. As the piezoelectric device is inside the vacuum chamber, only one electrical feedthrough is needed for operating the probe. The distance between reference electrode and sample can be varied within ±0.5 mm simply by applying a dc voltage to the piezoelectric device in addition to the ac oscillator-driving voltage, not requiring any mechanical driving and feedthrough provisions. The probe was tested studying the adsorption of O2 on Pt(111). The sensitivity for contact potential measurements is better than 0.1 mV at a response time of 10 sec.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.