Abstract

The performance of Red 4034 Perspex dosemeters from Harwell Technologies in routine photon dosimetry in mixed fields was evaluated. The dosemeters were irradiated in core grid positions of the Portuguese Research Reactor under a photon dose rate of the order of 10 4 Gy / h , a thermal neutron flux in the range of 1.8 × 10 10 to 1.1 × 10 11 n / cm 2 / s and temperatures below 40 ∘ C . The dosemeters visibly darken upon irradiation and this effect, accurately measurable by spectrophotometry, is related to absorbed dose. Thermoluminescence dosemeters and a calibrated ionization chamber were used as comparison. The results show that Red 4034 Perspex is a suitable dosemeter for mixed-field dosimetry up to 50 kGy .

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.