Abstract

The incidents at the Three Mile Island and Chernobyl have triggered the need for better personnel dosimetry methods in mixed radiation fields. This thesis presents a detailed computational study of a new method for mixed radiation field dosimetry using single-element TL dosimeters with pulsed laser heating schemes. The main objective of this study was to obtain an optimum heating scheme so that the depth-dose distribution in a thick TL dosimeter could be accurately determined. The major parts of the study include: (a) heat conduction calculations for TL dosimeters with various heating schemes, (b) glow curve calculations for TL dosimeters based on a first-order kinetic model, (c) unfolding of the depth-dose distribution based on the glow curve data, and (d) estimation of shallow and deep doses from the unfolded depth-dose distribution. Two optimum heating schemes were obtained in this study. The first one was obtained for a focused laser beam, and the second one was obtained for a uniform laser beam. Both heating schemes consist of two processes: top surface heating and bottom surface heating, and each process in turn consists of a sequence of laser pulses with various heating durations and power levels. Compared to the ``true`` depth-dose distribution obtained using Monte Carlo transport code EGS4, relative errors associated with the shallow and deep doses obtained from the unfolded depth-dose distributions are 5% and 25%, respectively, for the focused laser beam, and 15% in both doses for the uniform laser beam. 74 refs., 148 figs.