Abstract
In this work, the performance of a novel multi-pinhole SPECT (MPH SPECT) system based on Insight NM/CT Pro (Beijing Novel Medical Equipment Ltd) was evaluated. The sensitivity, spatial resolution, and quantitative accuracy of the MPH system were evaluated in conjunction with the methods in NEMA NU 1-2018. In addition, the performance of MPH in myocardial perfusion imaging was validated by phantom study and clinical study. The average sensitivity of MPH SPECT at the center, half, and edges of field of view (FOV) were 510 count s−1 MBq−1 (cps/MBq), 490 ± 85 cps/MBq and 540 ± 170 cps/MBq, respectively; the spatial resolution is about 6.5 ± 0.7 mm, 6.4 ± 0.8 mm, and 5.4 ± 0.4 mm for X, Y, and Z axes, respectively; the quantitative accuracy is superior to the conventional SPECT with low-energy high-resolution collimator (LEHR SPECT). The images of the phantom study and clinical study demonstrated the MPH SPECT can take only 5 minutes to obtain myocardial images with high image quality.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.