Abstract

Endoscopic retrograde cholangiopancreatography (ERCP) is a standard technique for the diagnosis and treatment of disorders of the pancreas or bile ducts. The aim of this study was the measurement of the radiation dose to patients during therapeutic ERCP procedures, in order to estimate the patient effective dose (ED). Fifteen patients were studied using a fluoroscopy system equipped with automatic brightness control and pulse fluoroscopy mode. Fluoroscopy time (FT), cumulative dose (Ka,r) and air kerma-area product (PKA) were collected for ERCP procedures. The ED was calculated from PKA values. The FT was ranged from 0.68 to 5.57 min, with the mean value of 2.50 min; the Ka,r was ranged from 2.22 to 19.10 mGy, with the mean value of 7.71 mGy; and the PKA was ranged between 0.59 and 5.10 Gycm2, with the mean value of 2.03 Gycm2. The ED ranged from 0.11 to 0.97 mSv, whilst the mean and median ED values were 0.39 and 0.32 mSv, respectively. FT and radiation dose to the patients were either comparative or significantly lower than those previously reported.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.