A novel large-area (20-inch) photomultiplier tube based on microchannel plate (MCP-PMTs) is proposed for the Jiangmen Underground Neutrino Observatory (JUNO) experiment. Its photoelectron collection efficiency Ce is limited by the MCP open area fraction (Aopen). This efficiency is studied as a function of the angular (θ), energy (E) distributions of electrons in the input charge cloud and the potential difference (U) between the PMT photocathode and the MCP input surface, considering secondary electron emission from the MCP input electrode. In CST Studio Suite, Finite Integral Technique and Monte Carlo method are combined to investigate the dependence of Ce on θ, E and U. Results predict that Ce can exceed Aopen, and are applied to optimize the structure and operational parameters of the 20-inch MCP-PMT prototype. Ce of the optimized MCP-PMT is expected to reach 81.2%. Finally, the reduction of the penetration depth of the MCP input electrode layer and the deposition of a high secondary electron yield material on the MCP are proposed to further optimize Ce.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.