Abstract

AbstractConditions for maximizing resolution and signal‐to‐noise ratios in fluorescence scanning densitometry of separations obtained by high performance thin‐layer chromatography are described. Resolution was found to be independent of most instrument parameters with a small decrease observed at wide slit widths. Signal was found to be markedly dependent on experimental parameters such as slit width, slit height, sensitivity setting, and the time constant of electronic filters. For recording chromatograms a slit width of 1.00 mm, a slit height slightly greater than the spot diameter, a low sensitivity setting (50 mV), electronic filtering, and a scan rate of 24 or 48 mm min−1 are commensurate with minimum recording time and maximum sample detectability without incurring a significant decrease in resolution.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.