We obtained a magnetic resonance image of 1 microm resolution and 75 microm(3) voxel volume for a phantom filled with hydrocarbon oil within an hour at 14.1 T. For this work, a specially designed probe with a high sensitivity RF coil and gradient coils generating over 1000 G/cm was built. The optimal pulse sequence was analyzed in consideration of the bandwidth, diffusion coefficients, and T(1) and T(2) relaxations of the medium. The system was applied to the in vivo imaging of a geranium leaf stem to get the images of 2 microm resolution and 200 microm(3) voxel volume.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.