Abstract

Ambient dose equivalent was determined in high energy reference radiation fields at CERN (CERF facility) using a recombination chamber and recombination methods developed in IAE. The chamber was also used for measuring the low LET background radiation which locally accompanies the fields at CERF. The measurements included determination of the absorbed dose and recombination index of radiation quality at different beam intensities. It was shown that the background might considerably influence the measurements of the absorbed dose, however, its influence on the ambient dose equivalent remains important only at low beam intensities.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.