ABSTRACTThe rapid electrochemical determination of Aceclofenac (ACF) has been employed by cyclic voltammetry (CV), differential pulse voltammetry (DPV) using developed OH-functionalised multiwalled carbon nanotube carbon paste electrode (OH-MWCNT/CPE). Modified electrode was characterised by scanning electron microscopy (SEM), energy dispersive X-ray spectroscopy (EDAX), X-ray diffraction spectroscopy (XRD), electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV). The ACF exhibits two oxidation peaks at +0.4 V, +0.66 V and one reduction peak at +0.3 V. The active surface area of the bare carbon paste electrode (BCPE) and modified electrode have been characterised by using K3[Fe(CN)6] solution containing 0.1 M KCl. In DPV mode, variation of ACF gave the limit of detection (LOD = 3s/m) 0.246 μM over the concentration range 1.0 to 190.0 μM (R2 = 0.9994). The developed electrode has good stability, reproducibility and could be successfully validated for the detection of ACF in pharmaceutical samples and biological fluids.