Continuous glucose monitoring (CGM) using transcutaneous sensors is becoming a sophisticated method to control and regulate glucose metabolism. The transcutaneous sensor of the CGM system (CGMS Medtronic Minimed, Northridge, CA, USA) is chosen to measure glucose concentration in interstitial fluid up to three days after insertion even though its function remains stable for a longer period. The question arises, which factors really limit the period of sensor insertion without unnecessary risk. The aim of this study was to assess any adverse events occurring in the course of 9 days after the sensor insertion. In a group of 22 healthy volunteers aged 21.8+/-1.30 y (mean +/- SE) a total of 26 sensors was inserted subcutaneously in gluteal or lumbar region for 9 days. Before insertion the site was sprayed with an antiseptic (Cutasept F, Bode Chemie, Hamburg, Germany). Local adverse reactions and disturbances in general condition were examined. In the course of 184 sensor-days, there were only minor local adverse events: hypersensitivity, itching, pain, redness, burning, subcutaneous hemorrhage. Additionally, sleep disturbances, attention deficits, problems related to the CGMS monitor, to adhesive tape and/or sensor were found. None of these resulted in sensor withdrawal. In 12 volunteers (55 %) no complications were observed. The sensor function measured according to electrical signals (ISIG) failed (always on day 1-2) in 4 cases (16 %). The present FDA approved 3-day insertion period for Medtronic transcutaneous sensor does not seem to limit its use and appears to be worth a careful revision.

Highlights

  • METHODSContinuous glucose monitoring (CGM) of glucose concentration in interstitial fluid is becoming a sophisticated method to control and regulate glucose metabolism[1, 2]

  • The transcutaneous CGMSTM sensor is labeled to measure the glucose concentration in interstitial fluid for up to three days after insertion even though it remains stable for a longer period[5,6,7,8,9,10,11]

  • There was no Adverse Event (AE) resulting in sensor removal

Read more

Summary

Methods

In a group of 22 healthy volunteers aged 21.8±1.30 y (mean ± SE) a total of 26 sensors was inserted subcutaneously in gluteal or lumbar region for 9 days. Local adverse reactions and disturbances in general condition were examined

Results
METHODS
RESULTS
DISCUSSION
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.