Abstract

We report new enzyme-containing siloxane membranes for biosensor elaboration. Lactate oxidase immobilization from water-organic mixtures with a high concentration of organic solvent (90%) leads to advanced lactate biosensors. The use of the new alkoxysilane monomers-(3-aminopropyl)trimethoxysilane (APTMS) and trimethoxy[3-(methylamino)propyl]silane (MAPS)-as the base for enzyme-containing membrane construction resulted in a biosensor with up to a two times higher sensitivity (0.5 A·M-1·cm-2) compared to the biosensor based on (3-aminopropyl)triethoxysilane (APTES) we reported previously. The validity of the elaborated lactate biosensor for blood serum analysis was shown using standard human serum samples. The developed lactate biosensors were validated through analysis of human blood serum.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call