Abstract

The measurement of local surface potentials by Kelvin force microscopy (KFM) can be sensitive to external perturbations which lead to artifacts such as strong dependences of experimental results (typically in a ∼1 V range) with KFM internal parameters (cantilever excitation frequency and/or the projection phase of the KFM feedback-loop). We analyze and demonstrate a correction of such effects on a KFM implementation in ambient air. Artifact-free KFM measurements, i.e., truly quantitative surface potential measurements, are obtained with a ∼30 mV accuracy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.