Abstract

Dielectric measurements using a probe consisting of a coaxial transmission line with an open-circuit end placed against the sample are discussed. For the 2.99- or 3.6-mm (OD) probes considered, a simple lumped parameter model shows errors above 1 GHz that increase greatly with frequency. An approximate model based on measured probe impedances from 1 to 18 GHz with samples consisting of water, methanol, and dioxane-water mixtures is evaluated. This model is more accurate than the lumped-parameter model and is better suited for calibration of the automatic network analyzer (ANA). The errors introduced in dielectric measurements by the use of approximate models for the probe are discussed. The technique succeeds because of partial cancellation of errors in modeling the probe in ANA-based measurements. >

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.