Abstract
New, simple, sensitive and rapid method for the determination of propranolol in the pharmaceutical preparations and human fluids. The Buildup and electrochemical response characteristics of a poly vinyl chloride (PVC) selective membrane electrode for the determination of propranolol (POP) are described in this research. The proposed electrodes was composed of propranolol-bromophenol blue as ion-exchanger and Di-butyl phthalate (DBPH) (electrode A), tris (2-ethylhexyl) phosphate (TEHP) (electrode B), and ortho-nitrophenyloctylether (ONPOE) (electrode C), as plasticizers. The slope was 50.94, 57.77, and 49.75 mV/decade for electrode A, B and C. the linear range was 5 ×10-5 – 1×10-2, 1×10-4 – 1×10-2, and 5×10-5 – 1×10-2 M, a detection limit of 4.2×10-5, 8.9×10-5, and 4.8×10-5 M, lifetime of 21, 42, and 1-day, respectively. Electrode B gives the best results, so the application of pharmaceutical and human fluids was based upon this electrode, the recovery percentage was 103, 101, 102.5, and 104 for POP tablet 10mg, tablet 40mg, urine and plasma by standard addition method, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.