Abstract
Controllable jet injection could provide a fast, repeatable method for delivery of small volume, high viscosity drugs to the dermis. We have developed a compact, lightweight jet injector that uses a small commercial rotary motor and differential screw to create a jet of fluid with sufficient force to deliver drug into tissue. Clear tissue analogue and ex vivo tissue were used to demonstrate repeatable delivery of <;10 μL of fluid using the device. The fluid when forced through a narrow orifice (108 μm) reaches a velocity of >250 m/s with an injection time of <;3 ms.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.