Abstract

For the simultaneous determination of Ni, Co, Cu and Cr by furnace atomic non-thermal excitation spectrometry (FANES), the optimum atomisation temperature Tat, gas discharge pressure pAr and discharge current iHK of the FANES emission source with respect to the signal to noise ratio (SNR) are determined. As the maximum values of each of the SNRs are not achieved simultaneously, a statistical optimisation method including both grid search and differential extreme value evaluation is applied to the compromise conditions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.