Abstract

A method of determining phosphorus by furnace atomic non-thermal excitation spectrometry (FANES) has been developed. The phosphorus is introduced into the FANES source as a solution of sodium dihydrogen phosphate. The measurements are carried out at 213.5/213.6 and 253.3/253.5 nm, the transition at 213.5/213.6 nm being the more sensitive one. The thermal and chemical conditions were optimised, the best chemical modifier being La3+ ions as they stabilise phosphate ions as LaPO4. Using the optimum amount of lanthanum (2 µg) the pre-treatment temperatures could be increased considerably (300 to 800 °C at normal pressure and 350 to 400 °C at low pressure). The best detection limit obtained is 90 pg of phosphorus, which is an improvement by a factor of 60 in comparison with the best values obtained with commercially available electrothermal atomisation atomic absorption spectrometric (ETAAS) instruments. Examples for real analyses with associated matrix interferences are given.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.