Abstract
A method for the determination of phosphorus by the molecular non-thermal excitation spectrometry (MONES) of PO and HPO has been developed. The phosphorus was introduced into the furnace atomic non-thermal excitation spectrometry (FANES) source as a solution of sodium dihydrogen phosphate. The measurements were carried out at the bands that gave the maximum signals, for PO at 246.4 nm and for HPO at 507 nm. The electrical, thermal and chemical conditions were optimised and are discussed. In addition, the influence of La3+ as a chemical modifier on the MONES of PO and HPO was investigated. The best values for the detection limit were 700 pg of P (MONES of PO) and 3400 pg of P (MONES of HPO). By comparison with the best values obtained with commercial electrothermal atomisation atomic absorption spectrometry instruments, these values are better by factors of 8 and 1.6, respectively. However, the determination of P by FANES at 213.5/213.6 nm is the best method for investigations of phosphorus using the FANES source, in the presence of La3+ as a chemical modifier.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.