Abstract

Isomer excitation by gamma,gamma' reactions and aluminium oxide thermoluminescence dosemeters (TLDs) have been used to monitor bremsstrahlung from the 4 MeV electron beam of a linear accelerator type LPR4 produced on a 0.9 mm Pt converter foil. Natural indium and osmium as well as TLDs have been irradiated at different distances (2-11 cm) and angles (0 degrees-90 degrees). Dose rates measured by TLDs were 5-110 kGy x h(-1). Isomer excitation of 115In (half-life 4.5 h) was used for monitoring bremsstrahlung of energies above 1 MeV, while that of 189Os (half-life 5.8 h) extended the available range down to 200 keV. Isomer production yields measured by gamma spectrometry and found to be about 10(-19)-10(-18) Bq per nucleus were calibrated against dose rate. A graphical method based on a semiempirical formula was used to evaluate the bremsstrahlung flux as well as the dose rate from the activity of isomeric monitors with uncertainties below 20%. The method is simple, of linear response in a large scale, independent of temperature, and able to monitor extremely high gamma intensities.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.