Abstract

We conceived new oscillometric blood pressure (BP) estimation methods based on physical modeling. The crux of these methods is to simultaneously estimate the arterial stiffness and BP of the patient from a standard oscillometric waveform. Hence, in contrast to previous methods, the BP estimation is specific to the patient at the time of measurement and is robust against arterial stiffening. We conducted initial testing of one of the methods against invasive reference brachial BP measurements in eight cardiac catheterization patients before and after nitroglycerin infusions. The method achieved systolic, diastolic and mean BP root-mean-squared-errors of 7.2, 7.6 and 6.7 mmHg. These errors were, on average, 40% lower than an existing high-end method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.