Abstract

An axisymmetric Wolter type mirror is used in a synchrotron radiation beamline to focus 150 eV soft x rays for microscopic x-ray photoelectron spectroscopy (XPS). The focusing characteristics are evaluated for x-ray source sizes of 1 mm, 200 and 120 μm in diameter. In this evaluation, the Si(2p) photoelectrons from a Si sample are counted while a knife edge is scanned across the focal point of the mirror. The lateral resolution is 2 μm by the 25%–75% criterion, and 4 μm by the full width at half-maximum definition in the beam intensity profile. The Wolter type mirror is expected to provide microscopic XPS techniques with micrometer to sub-micrometer resolution.