<p>Front cavity depth of microphone has influence in the determination of laboratory standard microphone sensitivity. Therefore this parameter is included in the measurement uncertainty budget. To determine microphone sensitivity accurately, it is necessary to measure the actual front cavity depth instead of using nominal value. This paper explains the measurement of LS1P and LS2P standard microphone front cavity depth using optical depth measurement facilities at SNSU-BSN and its effect on determining microphone sensitivity. The measurement was obtained from 4 positions distributed over the diaphragm for each microphone. The front cavity depth measurement result for LS1P is 1,94 mm ± 0,01 mm and for LS2P is 0,48 mm ± 0,01 mm These results comply with IEC 61094 Measurement Microphones-Part 2: Primary Method for Pressure Calibration of Laboratory Standard Microphones by Reciprocity Technique as the results are within the permissible range.</p>
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.