Abstract
The Institut National de Métrologie has set up an equipment for primary pressure calibration of standard condenser microphones by the reciprocity technique in a closed coupler.Although this calibration method is very classical and has been standardized in several countries, the accuracy of the measurement has been improved by a careful investigation of electrical and acoustical errors. A new method for the measurement of the acoustical impedance of microphones has been specially defined and developed. The uncertainties resulting from the equipment or from the measurement process have been greatly reduced owing to the high quality of the apparatus.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.