Abstract

A carbon fiber microcylinder electrode ( r = 3.5 μm) is used as a detector for high-efficiency and high-speed liquid chromatography. The microcylinder has a detection volume of a few picoliters, and can be placed directly at the outlet frit of the column. With proper positioning equipment, the electrode can be placed at the region of the outlet frit where the separation efficiency is highest. Such selective sampling results in greatly increased measured efficiency over a conventional electrochemical detector when short (4 cm) columns of conventional diameter (3.2 mm) are used. The microcylinder detector is sensitive and subpicomole detection limits are obtained in less than 30 s for norepinephrine. The need for expensive positioning equipment is eliminated by mounting the electrode into a fitting which can be mated directly with the column.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.