Abstract

A urinary glucose assay has been investigated, employing a micro flow injection analysis (microFIA) combined with a separation technique of glucose from the analyte. The adsorption part using activated alumina for the glucose in the analyte can be successively integrated onto a microFI chip. The selective adsorption-desorption of glucose in the artificial urine can progress on the adsorption part. Along with this selective preconcentration of glucose, the typical FI peak of glucose can be obtained just by feeding the sample and deionized water as an elutant sandwiched with the reagent on the carrier stream. The glucose concentration in artificial urine can be quantitatively determined with the present microFIA system, while the interference of other components coexisting in urine occurs in the case of the conventional FIA system without any separation part. The described method serves as a template for improving the selectivity for the analyte in the multi-component system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.