Abstract

Technical aspects of cardiac output determination by the indicatordilution method, using Cardio-Green and the densitometer 103 IR Gilford, are considered: these involve linearity of the device and influence of changing blood oxygen saturation, constancy of the amount of dye injected, surface integration of the curves and calibration technique.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.